Fluke Calibration 734C 4 Cell 10V DC Reference Std.
Tuotekoodi: 4992532
732C provides 10 V, 1 V, and 0.1 V outputs and is portable for interlab transfers or remote calibration needs.
734C includes four electrically and mechanically independe
Mikäli tuotteella ei ole näkyvissä hintaa, pyydä tarjous.
Ideal for Artifact Calibration
When paired with Fluke 742A-1 and 742A-10k Resistance Standards, a single 732C makes a robust and compact Artifact Calibration support package for instruments such as the Fluke 5730A High Performance Multifunction Calibrator and Fluke 8508A Reference Mulitmeters, including the older generation 5700A and 5720A models.
The most accurate and stable DC Voltage reference for demanding calibration environments
With the 734C, it is remarkably simple to establish and maintain a primary traceable voltage standard in your laboratory. Over time, with frequent intercomparisons of your four units, and regular calibrations of one or more units, you can achieve up to three times lower uncertainty with your 734C system.
From 1984 until the acquisition of our Josephson Array, the Fluke Calibration Primary Standards Laboratory maintained its corporate volt in this manner, reducing the absolute uncertainty to ± 0.35 μV/V — all traceable to national standards.
The 732C now includes 1 V and 0.1 V outputs
The 1 V and 0.1 V are key calibration and verification points for digital multimeters. Thanks to the high-precision thin-film resistive networks, which are manufactured in Fluke’s own thin-film fab facility, the 732C now comes standard with these additional precision outputs, eliminating the need for external dividers and making the measurement setup simpler and more reliable.
Select models for high-stability applications
Fluke Calibration offers Select models for customers requiring higher accuracy and lower drift for demanding workloads, or for labs that want greater independence from external calibration. The calibration process is the same for both Base and Select models; the difference is that the Select models (732C/S/C or 734C/S/04) provide two times better 10 V stability.
- Tuotekoodi: 4992532
| Performance Specifications | |||
| Output Voltages | 10 V, 1 V, and 0.1 V are provided at separate binding posts referenced to the VCOM binding post. | ||
| Stability | Stability for the 732C outputs at Tcal ±1 °C and IN CAL indicator on is specified in Table 2. Table 2. | ||
| Table 2. Standard Stability | |||
| Output Voltage | Stability (± µV/V) | ||
| 30 Days | 90 Days | 1 Year | |
| 10 V | 0.3 | 0.8 | 2.0 |
| 1 V | 0.6 | 1.2 | 3.0 |
| 0.1 V | 1.2 | 2.9 | 9.8 |
| Table 3. Select Stability | |||
| Output Voltage | Stability (± µV/V) | ||
| 30 Days | 90 Days | 1 Year | |
| 10 V | 0.3 | 0.8 | 1.0 |
| 1 V | 0.6 | 1.2 | 2.5 |
| 0.1 V | 1.2 | 2.9 | 8.0 |
| Noise at the Output Terminals | Output noise is specified for both day-to-day observations and for short-term observations, at k=1. | ||
| Table 4. Noise at the Output Terminals | |||
| Output Voltage | S1 (±µV/V) [1] | Sra (±µV/V) [2] | Noise (0.01 Hz to 10 Hz) (±µV/V rms) |
| 10V | 0.07 | 0.05 | 0.06 |
| 1 V | 0.16 | 0.14 | 0.15 |
| 0.1V | 1.4 | 1.3 | 1.0 |
| [1] S1 is the standard deviation about the regression (SDEV) of 90 days of at least twice daily stability test data. [2] Sra is the SDEV of the stability test data with a 7-day moving average filter (MAF). | |||
| To obtain the best performance, use the Product in controlled environments with good system grounding and shielding practices. For radiated EMI fields from 0.25 to 1 V/m from 80 to 130 MHz, add 9 µV to the 1 V output, and 3.6 µV to the 0.1 V output. For ac mains conducted EMI of 1 Vrms from 75 to 80 MHz, add 1 µV to the 1 V output, and 0.7 µV to the 0.1 V output. The 10 V output is mostly unaffected by EMI fields up to 1 V/m or by conducted EMI up to 1 Vrms. | |||
| Table 5. Output Current and Impedance | ||
| Output Voltage | Output Current Limit | Output Impedance |
| 10V | 12mA [1] | ≤1mΩ |
| 1V | 1.2mA [1] | ≤1mΩ |
| 100mV | 20pA | ≤100Ω |
| [1] Limit total output current to £0.1mA to realize specified battery operation | ||
| Table 6. Retrace (Hysteresis) Error | |
| Table 6 shows the change in 10V output voltage following a power outage (with the battery turned off) and temperature held constant in a 23 °C to 35 °C range | |
| Period that Power is Turned Off | Change in 10 V Output Value (±µV/V) |
| ≤10min | 0.1 |
| 10min to 24hr | 0.25 |
| 24hr to 14days | 0.25 |
| Stabilization Time Requirements | |
| Warmup times required after ac line and battery power have been turned off. The IN CAL indicator will be off, and recalibration will be necessary. The previously-specified retrace error specification can be used in the case of power interruptions. | |
| With no power interruption | No stabilization time is required after the Product is moved to another environment. |
| Power off for <1 hr | 1-hr warmup required |
| Power off for >1 hr | 24-hr warmup required |