Fluke Calibration 5550A Multi-Product Calibrator

Tuotekoodi: 5215654

Performance Multi-Product Calibrator

Mikäli tuotteella ei ole näkyvissä hintaa, pyydä tarjous.

5550A: Optimized for your multimeter workload

The 5550A Performance Multi-Product Calibrator enhances the 5522A Multi-Product Calibrator with new calibration ranges optimized to better match your existing digital multimeter workload in a modern, ruggedized instrument with an intuitive graphical interface. It calibrates the most popular benchtop digital multimeters with improved test uncertainty ratios (TURs), oscilloscopes to 1 GHz, and clamp meters to 1500 A with continuous current output. Technical improvements include 30 A continuous current output, which reduces wait time during heavy usage. A range redesign enables you to calibrate your existing benchtop multimeters with greater confidence. A larger variety of thermocouple types and a new thermocouple connector enables easier connection and broadens the temperature workload coverage. The workload is expanded even further when the 5550A is used with a 52120A amplifier to increase the current output to 6000 A, for calibrating high-current components such as Rogowski coils.

Key Features

Efficient, effective, productive -- your key to exceptional signal performance

  • The only commercially available multichannel oscilloscope calibration system
  • Simultaneous outputs on all channels increase tests speed and efficiency and eliminates the need for lead changes
  • Double bandwidth with a dedicated scope calibrator over a multi-product calibrator
  • More accurate signal to test a wide range of signals and frequencies
  • 4 channels plus trigger
  • Leveled sine waves up to 4 GHz
  • Expand your calibration abilities with performance and flexibility using Active Head Technology™ generating calibration signals right at the oscilloscope input
  • Fully automated with MET/CAL Software for hands-free operation

  • Tuotekoodi: 5215654

Specifications: Fluke 5560A, 5550A, 5540A and 5530A Multi-Product Calibrators

Target Workload 5560A 5550A 5540A 5530A
4.5 digit class multimeter, Example: Fluke 87V TUR 4:1 TUR 4:1 TUR 4:1 TUR 4:1
5.5 digit class multimeter, Example: Fluke 8808A TUR 4:1 TUR 4:1    
6.5 digit class multimeter, Example: Keysight 34460A TUR 4:1      
Best Performance @95%, ±5°C, 1 year Range
DCV 0 V to ±1020 V 6 ppm 9 ppm 24 ppm 24 ppm
DCI 0 A to ±30.2 A continuous 78 ppm 85 ppm 100 ppm 100 ppm
ACV 1 mV to 1020 V; 3 Hz to 500 kHz 110 ppm 120 ppm 780 ppm 780 ppm
ACI 10 uA to 30.2 A continuous; 3 Hz to 30 kHz 120 ppm 390 ppm 1600 ppm 1600 ppm
Variable Resistance 0 Ohm to 1200 MOhm 19 ppm 22 ppm 50 ppm 50 ppm
Variable Capacitance 0 pF to 120 mF 0.09% 0.16% 0.19% 0.19%
Variable Inductance 13 uH to 120 H 0.09% 0.16% - -
Temperature 17 TCs 10 RTDs; ITS-90 or IPTS-68 0.09°C TC; 0.002°C RTD 0.09°C TC; 0.002°C RTD 0.09°C TC; 0.002°C RTD 0.09°C TC; 0.002°C RTD
Functional Differences
AC Power (Phase Adj. and PF); Dual Output (DCV-DCV, ACV-ACV) Yes Yes - Yes
Multi-Unit Sync Yes Yes Yes -
Simple Harmonics Yes Yes - Yes
Inductance Yes Yes - -
52120A Control (Closed Loop); Up to 3 or 360 A Yes Yes Yes Yes
Fluke Coil Effective Current Display Yes Yes - -
Extended Volts/Hertz ACV - 70 V at 300 kHz Yes - - -
Specification Period, 95% and 99% Confidence 90 day; 1 year; 2 year 90 day; 1 year 1 year 1 year
Fast Settling Time Yes Yes Yes Yes
Isolated terminal block (DMM Autocal Compatible) Yes Yes Yes Yes
Scope Output Options
Leveled Sine Bandwidth; < 4% @ 2 GHz 600 MHz; 1 GHz; 2 GHz 600 MHz; 1 GHz 600 MHz -
Specified VSWR Yes Yes Yes -
Vertical Amplitude (AC & DC) 0 V to 120 V 0 V to 120 V 0 V to 120 V -
Fast Edge < 175 ps < 175 ps < 175 ps -
Time Marker Accuracy 2.5 ppm 2.5 ppm 2.5 ppm -
Pulse Yes Yes Yes -
Impedance Measurement Yes Yes Yes -
Input Overload Yes Yes Yes -

 